SOP for Mechanical Qualification of Dissolution Apparatus 1 and 2

Recommended standard operating procedure ( SOP ) for Mechanical Qualification of Dissolution Apparatus 1 and 2

1. Purpose The purpose of this SOP is to establish the setup, mechanical calibration, and operation checks for dissolution Apparatus 1 (Basket) and 2 (Paddle).
2. Scope/Policy
This procedure applies to all Apparatus 1 and 2 dissolution equipment

3. Responsibilities
3.1 Analyst
• Check the vessel, basket, and paddle dimensions on receipt.
• Perform the maintenance procedures as per the manufacturer’s recommendation.
• Perform the mechanical calibration on receipt, after the instrument is moved, after the instrument is repaired, and six months after the previous calibration. If the instrument is not being used routinely the six month mechanical calibration can be performed before performing the first dissolution test after the six month time interval.
• Perform the operation checks at each time of use.

4. Background
The setup, mechanical, and operational checks are used to minimize variability during dissolution testing.
5. References
• USP General Chapter 711
• The Dissolution Procedure SOP

6. Procedure
Wherever possible, tools should be traceable .
6.1 Apparatus setup
During apparatus setup or after replacement of parts, verify the following dimensions. Certificates of Analysis (COA) or Certificates of Conformity (COC) may be used to document the measurements. Discard any parts that do not meet specifications.
Vessel Dimensions
Use an appropriate measuring device to verify that the vessel dimensions conform to the specifications listed in the USP General Chapter 711 Dissolution. The vessel must have cylindrical sides and a hemispherical bottom which must be smooth and without defects.
Basket Dimensions
Each basket must conform to the dimensions shown in the USP General Chapter 711 Dissolution in Figure 1, Basket Stirring Element. An appropriate measuring device is used to make the measurements.
Paddle Dimensions
Each paddle must conform to the dimensions shown in the USP General Chapter 711 Dissolution in Figure 2. An appropriate measuring device is used to make the measurements.
6.2 Maintenance
Follow the manufacturer’s maintenance recommendations and maintenance schedule.

6.3 Mechanical Calibration
Perform the following in the order given for mechanical calibration of each apparatus. Some iteration may be necessary if extensive adjustments are needed. Perform these tests every 6 months or after any repair, move, etc. If the instrument is not in routine use, the mechanical calibration may be performed prior to performing the first dissolution test after the six month interval. Some dissolution instruments require the use of manufacturer’s supplied special tools or incorporate automatic mechanical devices to perform the following tests. These may be used provided they follow the general principle of the procedure.
Shaft Wobble
A runout gauge is placed on top of the vessel plate, and the drive module is positioned so that the gauge probe touches the shaft about 2 cm above the top of the paddle blade or basket. The gauge is placed so that the probe slightly presses in on the turning shaft. If a mechanical gauge is used, the gauge’s pointer should read slightly more than zero. The pointer will vary from a minimum to a maximum reading, and the difference is called the wobble. The specification is ≤1.0 mm total runout.
Paddle and Basket Shaft Verticality
Lower the drive unit to where it would be during an actual dissolution test. If necessary the shaft verticality may be checked with the shafts raised above the drive unit. Place an accurate bubble level on the front edge of each of the shafts. The bubble should be within the lines of the level. Rotate the level 90ยบ so it is on the side of the shaft. The bubble should again be within the lines of the level for each shaft. If the shafts are not vertical adjust the feet of the apparatus until they are vertical.
A digital leveling device may also be used to determine the shaft verticality. The shaft must be ≤0.5° from vertical.
Basket Wobble
A runout gauge is placed on top of the vessel plate and the drive unit is positioned so that the gauge probe touches the bottom rim of the basket. The gauge is placed so that the probe slightly presses in on the turning shaft. If a mechanical gauge is used, the gauge’s pointer should read slightly more than zero. The pointer will vary from a minimum to a maximum reading and the difference is called the wobble. The specification is ≤ 1.0 mm total runout.
Vessel Centering
The vessel plate of an apparatus may warp or bend or the thickness of a vessel lip or centering collar lip may not be perfectly uniform. If either of these occurs, even though the shafts are vertical and the vessel plate is level, the vessel walls may not be vertical.
Use centering tools which measure centering inside the vessel. Two centering tools are used to center the vessels around the paddle or basket shafts and to align the vessels so that their sides are vertical. For the paddle method, the bottom of one centering tool is placed 2 mm above the top of the paddle blade and the bottom of the second centering tool is clamped on the shaft 80 mm above the blade with the probes positioned in the same direction towards the glass vessel wall. For the basket method, the bottom of one centering tool is placed 2 mm above the top of the basket and the bottom of one centering tool is placed 60 mm above the top of the basket with the probes positioned in the same direction towards the glass vessel wall. Carefully lower the shaft and centering tools into the vessels so that the paddle blade or basket bottom is about 2.5 cm above the bottom of the vessel. Manually rotate the shaft slowly and check the centering at both levels. If the vessel is not centered at either level, adjust the vessel to center it. Adjustments can be made by rotating the vessel or the vessel with the centering collar inside the vessel plate, moving the vessel sideways within the vessel plate or placing shims (such as tape) under one side of the lip of the vessel or vessel centering collar. Repeat this process until both bottom and top positions are centered within 1.0 mm from the center line.
An alternative procedure is to use a mechanical or digital centering device that centers the inside of the vessel around the shaft or a surrogate shaft. The centering is measured at two positions inside the vessel in the cylindrical portion, one near the top but below the rim and one just above the bottom portion of the vessel. The shaft or surrogate shaft must be centered within 1.0 mm from the center line.
Vessel Verticality
The vessel verticality can be calculated using the centering measurements and the difference in height between the two measurements or it can be determined using a digital leveling device placed on the inside wall of the vessel. The verticality should be determined at two positions 90° apart. Adjustments can be made by placing shims (such as tape) under one side of the lip of the vessel or vessel centering collar. The vessel must be ≤ 1.0° from vertical.
After each vessel has been centered and made vertical, each vessel and vessel plate opening must be numbered and a mark must be placed on the lip of each vessel and on the vessel plate directly next to the mark on the vessel lip. Each vessel must be returned to the same vessel plate opening and positioned in the exact same position inside the vessel plate opening for all future dissolution tests.
Basket and Paddle Depth
The actual distance between the bottom of the vessel and bottom of the basket or paddle is determined. If the depth of the basket/paddle is adjustable, first a depth gauge is used to set the distance between the bottom of the paddle blade or basket and the bottom of the vessel. The depth gauge is set at 25 mm and placed on the bottom of the vessel. Each shaft is raised into the apparatus drive module. The drive unit is then lowered to its operating position. The paddle or basket is then lowered into the vessel until it touches the top of the depth gauge. The shafts are locked into this height. This is repeated for each shaft. The specification is 25 mm ± 2 mm.
Rotational Speed
A tachometer should be used to measure the rotational speed of the paddle or basket. The shafts should be rotating smoothly at ± 2 rpm of the target value.
6.4 Operation
Before each test perform the following:
Basket Examination
Each basket must be visually examined for defects such as rusting, corrosion, wires sticking out beyond the basket, clogged mesh holes or deformed mesh sides.
Paddle Examination
Each paddle must be visually examined for defects such as rusting, corrosion or loose pieces of coating on the paddles (for paddles coated with Teflon or another coating).
Vessel Examination
Each vessel must be free of scratches, cracks, pits and residue.
Vessel Temperature
The temperature of the medium inside each vessel is measured at time of use. The limit is ± 0.5°C of the target temperature. The target temperature is usually 37°C for Apparatus 1 and 2.
Vibration
The USP criteria of: “No part of the assembly, including the environment in which the assembly is placed, contributes significant motion, agitation, or vibration beyond that due to the smoothly rotating stirring element” is followed.
6.5 Additional Variables
Basket Shafts (Clips versus O rings)
The diagram of the basket stirring element in the USP General Chapter 711 shows that the basket shaft has clips to hold the basket. Some basket shafts have O rings to hold the basket in place instead of clips. The clips change the hydrodynamics of the medium causing slightly increased dissolution results with certain formulations. In order to conform to the USP diagram, DPA chemists must use basket shafts with clips or use detachable basket clips unless the dissolution method states otherwise.
Sinkers
Sinkers are required for capsules that float when the Apparatus 2 (Paddle) method is used. Some commercial sinkers have too many coils that trap the capsule material inside the sinker.
uses sinkers that are recommended by the USP. A detailed procedure on how to make them is in the proposed USP General Chapter 1092 The Dissolution Procedure: Development and Validation, USP 31(5), 2005, p. 1463. If a method states to use a particular commercially available sinker, one must use the specified sinker.
7. Records
The date, analyst, dissolution vessels’ manufacturer, and the dissolution apparatus's manufacturer, model number, and serial number will be recorded on the appropriate Mechanical Calibration Report Sheet (see Photo) along with the appropriate observations. The completed report sheet will be placed in the report sheet folder for that apparatus. Each dissolution apparatus will have its own report sheet folder.
8.Report Format


=====================================================================================================
Do you know now this website has become a most popular and most referred website in pharmaceutical industry and pharmaceutical companies and pharmaceutical manufacturers from all over the world ,for almost all topics related to Pharmaceutical Manufacturing , Pharmaceutical Regulatory Affairs and Good Manufacturing Practice for Pharmaceutical Manufacturing (c GMP guidelines ) and latest news and new drugs developments.

=====================================================================================================
You may also like following articles

Whar is Referance Listed Drug  ? ( RLD )

What is Pharmaceutical Equivalents

What is Pharmaceutical Alternatives

What is Therapeutic Equivalents


Do Physical properties contribute to drug activity.

What is drug receptor , How a drug resistance occurs

Mechanism of drug resistance

What is drug interaction

Drug interaction, and its examples

What is first pass metabolism of a drug

What is What is 510(k) Clearances ?

What is 510(k) Clearances,

Premarket Notification for medical devices - PMN or 510(k)

What is a drug interaction

Examples of drug interactions


Antibiotic Definition and classification


Antibiotic resitance and Antibiotic resistance mechanism

Antioxidants food suppliments

Vitamin D Details on FDA cautions on accurate dosage of Vitamin D


What is an antibody? what is monoclonal and polyclonal antibodies?

Terminologies In vaccine Production

Multi stage testing of Virus vaccine production

Testing of vaccines at different stages of production

TESTING FOR ADVENTITIOUS AGENTS CELL PROPERTIES IN VIRAL VACCINE PRODUCTION

Enzyme linked immunosorbent assay ELISA

Raido Immuno assay

http://whoguideline.blogspot.com/2010/04/terminalogy-and-their-explanations.html

Pharmaceutical Aseptic Manufacturing Process Terms , Terminology and Definitions.

http://whoguideline.blogspot.com/2010/02/pharmaceutical-aseptic-manufacturing.html

Here are some articles which will be useful for you in further understanding of aspects of sterile dosage form manufacturing and regulatory affairs and good manufacturing practice in

pharmaceutical industry

Pharmaceutical Validation

Types of validations in pharmaceutical manufacturing


Requirements of documents for validation of sterilisation process

How to investigate OOS out of specification results

Determination of Phenol coefficient of a disinfectant

Sterility testing

Cleen Room Classification

Time limitations in sterile pharmaceuticals processing

Aspects of validation of manufacturing process in sterile pharmaceuticals

Clinical Trials

Controlling Pyrogens in injectable dosage forms

Media fill run process simulation aspects Validation of Aseptic Process and Sterilisation

New Drug Application (NDA) how to make a New Drug Application (NDA) to US FDA

Abbreviated New Drug Application (ANDA) What is ANDA , detaied information about ANDA preparation and submission to US FDA

How to make Investigational New Drug (IND) Application to US FDA

Drug applications submission to us fda Over the counter Drugs OTC drugs

BIOAVAILABILITY AND BIOEQUIVALENCE REQUIREMENTS

Electronic record in pharmaceutical manufacturing industry

Standard Operating Procedure

Good manufacturing practice in pharmaceutical industry

Pharmaceutical industry pharmaceutical companies and FDA latest updates

Here is an interesting article on world wide pharmaceutical industry and making a carrier in one of pharmaceutical companies article on Pharmaceutical Industry pharmaceutical industry

Find a Job in Pharmaceutical Company
To Find Pharmaceutical jobs and make a Pharmaceutical careers see here pharmaceutical companies
Here are some interesting articles on Quality assurance systems for pharmaceutical company

1.Quality assurance in pharma industry

2.Quality by designe concept for pharmaceutical industry

3.Quality by designe concept in pharmaceutical industryan explanation

No comments: